|
Safety with the
Electrosurgery Generator
Protection Against
Low Frequency Electric Shocks
The "electric shock",
associated with the domestic mains supply, is caused by the stimulating
nature of the low frequency alternating current (commonly 50 Hertz). The
electrosurgery generator, by use of special transformers and optical isolating
devices, ensures that dangerous levels of the low frequency current cannot
reach the patient or operator. Cardiac arrest, fibrillation or respiratory
arrest could result from malfunctioning equipment without such isolation.
Most generators cannot
detect bad or loose connections in the active electrode, which could result
in higher than normal low frequency components in the electrosurgery current.
Low frequency current components can cause fibrillation, and it is advised
that the patient plate (dispersive electrode) is placed so that the electrosurgery
current will not pass directly through the region of the heart.
Safety Guidelines
in the Use of Electrosurgery
Routine safety checks
are essential - read and learn the Do's and Don't's for the correct use
of electrosurgical generator and associated accessories:
DO's
- Place the patient
plate as close to the surgical site as possible, with as much smooth
surface contact as possible.
- Place any monitoring
electrodes (i.e. ECG) as far as possible from the active electrode to
patient plate electrosurgical current path.
- Locate the patient
plate over well vascularised muscular areas.
- Avoid unnecessary
or prolonged activation of the electrosurgical unit.
- Keep active electrode
tips clean and free of debris during use. Eschar or carbon adhered to
the electrode prevents proper current flow necessitating high power
settings.
- Check patient plate
and all connections during surgery if constant power level increases
are required.
- Read the Instruction
Manual thoroughly before using the equipment for the first time.
- Schedule routine
preventive maintenance of all electrosurgical equipment, and follow
equipment "Systems Check" on a regular basis.
DON'T's
- Never use electrosurgery
in the presence of explosive or flammable anesthetics, fluids or gases.
Normal electrosurgery produces sparking at the active electrode, which
could ignite those gases or liquids.
- Do not use uninsulated
active accessories, or instruments on which the insulated coating has
been broken. Use only insulated forceps, endoscopic sleeves and eyepieces.
Cords and cables which have broken, split or in any way damaged must
be replaced, not repaired.
- Don't re-use disposable
accessories. These can seldom be properly cleaned.
- Don't locate patient
plate over bony or hairy areas, scar tissues or implanted prostheses.
A bad contact could result and cause patient burning.
- Never allow pools
of any fluid to accumulate at either active electrode or patient plate.
This can cause unnecessary tissue destruction or patient burning.
- Don't cut the patient
plate to reduce the size at any time.
- Don't use unmonitored
plates on electrosurgical equipment with monitoring facilities. Protection
given by the machine would be lost.
- Don't use bent,
crumpled or torn patient plates, or any plate which in any way has suspect
conductivity or patient contact.
- Do not allow any
cables to become knotted.
- Never "test" the
output of the machine by touching the active electrode to ground.
- Do not place the
active electrode on a grounded surface when not in use.
- Never contact two
active monopolar forceps simultaneously.
|